Leybold PHOENIX Quadro Helium Leak Detector | 180° magnetic sector mass spectrometer, vacuum mode ≤ 5×10⁻¹² mbar·l/s, touch screen, HDMI, 41 kg
LEYBOLDPHOENIX Quadro
[ICANSTron] Genuine authorized LEYBOLD product

Leybold PHOENIX Quadro Helium Leak Detector | 180° magnetic sector mass spectrometer, vacuum mode ≤ 5×10⁻¹² mbar·l/s, touch screen, HDMI, 41 kg

Leybold PHOENIX Quadro Helium Mass Spectrometer Leak Detector: Vacuum Mode ≤ 5×10⁻¹² mbar·l/s Overview The PHOENIX Quadro is a helium mass spectrometer leak detector in the Leybold PHOENIX series. It measures helium (and also hydrogen) with a 180° magnetic sector mass spectrometer and supports vacuum mode and sniffer mode (requires a sniffer line).

Minimum detectable helium leak rate: vacuum mode ≤ 5×10⁻¹², sniffer mode < 1×10⁻⁹ mbar·l/s180° magnetic sector mass spectrometer; detectable masses 2, 3, 4 amu; start-up ≤ 110 sPumping speed 2.5 / 3.0 m³/h (50 / 60 Hz); 495 × 475 × 318 mm; 41 kg
Corporate reference price (excl. tax)
Price on request
E-invoice issued after payment
One-year manufacturer warranty
Chat with Us on LINE

Overview

Leybold PHOENIX Quadro Helium Mass Spectrometer Leak Detector: Vacuum Mode ≤ 5×10⁻¹² mbar·l/s

Overview

The PHOENIX Quadro is a helium mass spectrometer leak detector in the Leybold PHOENIX series. It measures helium (and also hydrogen) with a 180° magnetic sector mass spectrometer and supports vacuum mode and sniffer mode (requires a sniffer line). The manufacturer positions it for industrial production lines for analytical instruments, medical devices, food packaging, automotive and aerospace, as well as semiconductor and flat panel display equipment. It has a touch screen, can store multiple parameter sets and copy them to other PHOENIX leak detectors, and has an HDMI interface for an external monitor plus a USB port. The PHOENIX Quadro Dry Plus with a dry backing pump (SCROLLVAC 3 Plus) and the modular PHOENIX Vario are available on request.

Key Specifications

  • Minimum detectable helium leak rate: Vacuum mode ≤ 5×10⁻¹² mbar·l/s; sniffer mode < 1×10⁻⁹ mbar·l/s (depending on sniffer line and environment)
  • Minimum detectable hydrogen leak rate: Vacuum mode ≤ 1×10⁻⁸ mbar·l/s; sniffer mode < 1×10⁻⁷ mbar·l/s
  • Leak rate display range: 1×10⁻¹² to 1×10⁻¹ mbar·l/s
  • Maximum inlet pressure: 15 mbar; 1000 mbar with partial flow pump set
  • Pumping speed: Evacuation 2.5 m³/h (50 Hz), 3.0 m³/h (60 Hz); helium pumping speed in vacuum mode FINE 1.2 l/s, ULTRA over 3.1 l/s
  • Response and start-up: Leak signal time constant < 1 s; start-up ≤ 110 s
  • Mass spectrometer: 180° magnetic sector; 2 long-life yttrium / iridium cathodes; detectable masses 2, 3, 4 amu
  • Connections: Inlet DN 25 ISO-KF; backing pump connection 2 × DN 25 ISO-KF
  • Power supply: EU 230 V 50 / 60 Hz (250000V02); US 115 V 60 Hz (251000V02); JP 100 V 50 / 60 Hz (251100V02)
  • Dimensions and weight: 495 × 475 × 318 mm; 41 kg

Applications

  • Helium leak testing of vacuum chambers, piping and components.
  • Leak checks and maintenance of semiconductor and flat panel display equipment.

Specifications per the Leybold PHOENIX series catalog 3610 0460 02 Technical data (leybold.com, retrieved 2026-10-24).

Specifications

Key specifications
Minimum detectable helium leak rate: vacuum mode ≤ 5×10⁻¹², sniffer mode < 1×10⁻⁹ mbar·l/s180° magnetic sector mass spectrometer; detectable masses 2, 3, 4 amu; start-up ≤ 110 sPumping speed 2.5 / 3.0 m³/h (50 / 60 Hz); 495 × 475 × 318 mm; 41 kg
Model
PHOENIX Quadro
Brand
LEYBOLD
Purchasing
In stock, tax incl. / B2B formal quotation request (RFQ)
Warranty
One-year manufacturer warranty

Calibration

If your instrument needs periodic calibration, register the brand, model and calibration requirements on the calibration service page and we will send a separate quote.

Go to calibration service